Development of electrochemical aptasensor for lung cancer diagnostics in human blood A. V. Shabalina, D. O. Sharko, Y. E. Glazyrin [et al.]

Contributor(s): Shabalina, Anastasiia V | Sharko, Darya O | Glazyrin, Yury E | Bolshevich, Elena A | Dubinina, Oksana V | Kim, Anastasiia M | Veprintsev, Dmitry V | Lapin, Ivan N | Zamay, Galina S | Krat, Alexey V | Zamay, Sergey S | Svetlichnyi, Valerii A | Kichkailo, Anna S | Berezovski, Maxim VMaterial type: ArticleArticleContent type: Текст Media type: электронный Subject(s): электрохимические аптасенсоры | биосенсорный слой | рак легких | кровьGenre/Form: статьи в журналах Online resources: Click here to access online In: Sensors Vol. 21, № 23. P. 7851 (1-20)Abstract: We describe the preparation and characterization of an aptamer-based electrochemical sensor to lung cancer tumor markers in human blood. The highly reproducible aptamer sensing layer with a high density (up to 70% coverage) on the gold electrode was made. Electrochemical methods and confocal laser scanning microscopy were used to study the stability of the aptamer layer structure and binding ability. A new blocking agent, a thiolated oligonucleotide with an unrelated sequence, was applied to fill the aptamer layer’s defects. Electrochemical aptasensor signal processing was enhanced using deep learning and computer simulation of the experimental data array. It was found that the combinations (coupled and tripled) of cyclic voltammogram features allowed for distinguishing between the samples from lung cancer patients and healthy candidates with a mean accuracy of 0.73. The capacitive component from the non-Faradic electrochemical impedance spectroscopy data indicated the tumor marker’s presence in a sample. These findings allowed for the creation of highly informative aptasensors for early lung cancer diagnostics.
Tags from this library: No tags from this library for this title. Log in to add tags.
No physical items for this record

Библиогр.: 67 назв.

We describe the preparation and characterization of an aptamer-based electrochemical sensor to lung cancer tumor markers in human blood. The highly reproducible aptamer sensing layer with a high density (up to 70% coverage) on the gold electrode was made. Electrochemical methods and confocal laser scanning microscopy were used to study the stability of the aptamer layer structure and binding ability. A new blocking agent, a thiolated oligonucleotide with an unrelated sequence, was applied to fill the aptamer layer’s defects. Electrochemical aptasensor signal processing was enhanced using deep learning and computer simulation of the experimental data array. It was found that the combinations (coupled and tripled) of cyclic voltammogram features allowed for distinguishing between the samples from lung cancer patients and healthy candidates with a mean accuracy of 0.73. The capacitive component from the non-Faradic electrochemical impedance spectroscopy data indicated the tumor marker’s presence in a sample. These findings allowed for the creation of highly informative aptasensors for early lung cancer diagnostics.

There are no comments on this title.

to post a comment.
Share