Measurement of the energy resolution and calibration of hybrid pixel detectors with GaAs:Cr sensor and Timepix readout chip A. Butler, P. Butler, S. Bell [et.al.]

Contributor(s): Butler, A | Bell, Steven J | Chelkov, G. A | Dedovich, D. V | Demichev, M. A | Elkin, V. G | Gostkin, M. I | Kotov, S. A | Kozhevnikov, D. A | Kruchonak, U. G | Butler, P | Nozdrin, A. A | Porokhovoy, S. Yu | Potrap, I. N | Smolyanskiy, P. I | Zakhvatkin, M. M | Zhemchugov, A. S | Томский государственный университет Радиофизический факультет Научные подразделения РФФMaterial type: ArticleArticleSubject(s): Timepix, детектор | арсенид галлия | калибровкаGenre/Form: статьи в журналах Online resources: Click here to access online In: Physics of particles and nuclei letters Vol. 12, № 1. P. 59-73Abstract: This paper describes an iterative method of per-pixel energy calibration of hybrid pixel detectors with GaAs:Cr sensor and Timepix readout chip. A convolution of precisely measured spectra of characteristic X-rays of different metals with the resolution and the efficiency of the pixel detector is used for the calibration. The energy resolution of the detector is also measured during the calibration. The use of per-pixel calibration allows to achieve a good energy resolution of the Timepix detector with GaAs:Cr sensor: 8 and 13% at 60 and 20 keV, respectively.
Tags from this library: No tags from this library for this title. Log in to add tags.
No physical items for this record

Библиогр.: 12 назв.

This paper describes an iterative method of per-pixel energy calibration of hybrid pixel detectors with GaAs:Cr sensor and Timepix readout chip. A convolution of precisely measured spectra of characteristic X-rays of different metals with the resolution and the efficiency of the pixel detector is used for the calibration. The energy resolution of the detector is also measured during the calibration. The use of per-pixel calibration allows to achieve a good energy resolution of the Timepix detector with GaAs:Cr sensor: 8 and 13% at 60 and 20 keV, respectively.

There are no comments on this title.

to post a comment.
Share